DentalQC › Docs › Measuring page (Measure)

Measuring page (Measure)

The main desk: the left rail takes files and settings, the viewport is in the middle, the QC board on the right, the measurement tables at the bottom. The ? next to each block title opens a short explanation.

Measuring page screenshot
Measuring page screenshotThe icon buttons on the top bar open the Dashboard, Case library, comment board, comparison, batch measuring, Re-analyse and Export report.

Top bar

  • Score strip (after measuring): overall score /100, a four-zone colour gauge (below 60 red · 60–84 amber · 84–92 light green · 92+ green), the A–E grade with its meaning, and the “scored on x/y groups” line. Hover that line to see which groups counted and which did not.
  • Icon buttons Dashboard and Case library: leave the measuring page in the same tab.
  • Comment on a case button: opens the comment board in a new tab; there, click Get measured case to bring over the views you just measured.
  • Compare versions button: opens the comparison page in a new tab, carrying the case you just measured as the after version.
  • Batch measure button: opens the multi-case page in a new tab.
  • Re-analyse button: re-measures the open case with the current settings. Dimmed until a file is chosen or a result exists.
  • Export report button: opens the A4 report of the finished measurement in a new tab. Dimmed until a measurement finishes.
  • Language flags: switch in place without losing results.

Left rail › Files to measure

  • File type switch HTML webview / STL. Drop the wrong type and the app switches mode itself and tells you.
  • Case count switch Single case / Batch ↗: Batch opens the batch page with the file type currently selected.
  • Drop zone: drag the HTML file (HTML mode) or the case folder / .stl files with the .constructionInfo (STL mode) here. Dropping anywhere on the page works too.
  • Choose file opens the file picker; in STL mode there is also Choose case folder to pick a whole folder.
  • Under the drop zone the chosen file name appears (“— not measured” until you click Measure), and after measuring a case card: software, tooth count (upper/lower), gingiva/scan, margin source, tolerance profile.
  • The HTML webview already carries the 3D mesh: no screenshots and no scale calibration needed. In DentalCAD: Export → Web preview (HTML).

Left rail › Prescription (station 01)

  • Prescription box: paste the dentist's email, order form or chat message verbatim. What the parser read appears right below under Read from the sheet, and only what is shown there is used.
  • When the tooth numbering system is unclear, the FDI / Universal / Palmer question appears with the reason; the cross-check waits until you click one.
  • The Drop .constructionInfo button picks the exocad XML of the case; its name appears next to it and × removes it. The readout below shows how many teeth the file declares, MinThickness, MillingDiameter, cement gap, and whether the tooth set disagrees with the design.
  • Details of how prescriptions are read: see Prescription & .constructionInfo.

Left rail › Scope

  • Arch: Automatic (read both if present) · Maxillary only · Mandibular only.
  • Scored items: in QC protocol → Scoring thresholds, the Scored column switches an item off or on. A switched-off item is still measured and shown, but no longer takes points off; export checks never change. On personal and Free plans the choice is yours only; on the Lab plan the owner or an admin sets it for the whole lab. After your first measurement, a pop-up lists preference items (symmetry, esthetics, form…) that cost points and lets you switch them off and re-score at once — tick Don't show this again to stop it.
  • Tolerance profile: CAD/CAM lab — tight · Clinical — standard · Perceptual threshold — relaxed. The preselected value is the workspace default (set on Scoring thresholds).
  • The Your thresholds / Workshop thresholds button (free plan: Scoring thresholds) right under the profile box: opens the table of measurements scored with custom thresholds; the small number on the button is how many were changed.
  • Projection resolution slider: 10–26 px/mm, default 16. Higher is sharper but slower.
  • Viewing angles: Standard — 4 views (front, occlusal, right, left) · Full — 7 views (adds 45° three-quarter, lingual, along the insertion axis). The extra three are for looking only; every measurement still runs on the front view.

Left rail › Material, Tool, Patient

  • Restorative material (station 03): if undeclared, wall thickness is still measured and shown but not scored, and the thickness/connector gates read “not checkable”.
  • Preparation geometry (station 02): when the file carries the preparation scan of a tooth, DentalQC measures the preparation itself — taper (TOC, also split mesial-distal and buccal-lingual), wall height, height-to-width ratio, finish-line width and type, undercut along the path of insertion, and occlusal clearance to the opposing arch, with a polar chart of the taper per sector. The preparation is the dentist's work, so this group is shown but never counted in the overall score. Open the preparation report prints a one-page sheet for the clinic (and can be sent like the QC report).
  • Cement gap (station 03): when the exocad file carries both the design bottom and the preparation of a tooth, the gap between them is measured by zone — the seal band at the margin, the axial walls and the occlusal part — and shown with a colour map seen from the margin side. An intaglio that cuts into the preparation is scored and can block the case (gate Seats on the preparation). Drop the .constructionInfo in to compare the axial gap with the cement-gap value set in the design.
  • Implant restorations (station 03): from the exocad layers (analog, Ti-base, screw channel, abutment) DentalQC measures the screw channel angle to the implant axis, the thinnest wall around the channel, screw-head clearance, the bonding gap and bonding height on the Ti-base, the emergence angle and bulge around the implant axis, a custom abutment's height, taper and angle, and the divergence between splinted implants. Channel wall, channel angle, cut-in and bonding gap are scored (group Implant restoration) and checked by the gate of the same name; emergence and screw-access position are reference marks, shown but not counted.
  • Smallest bur (station 07): the smallest bur diameter of your mill. If not declared, concave radii are measured but not judged.
  • Bur reach on the intaglio (station 07): DentalQC runs a ball-end bur along the insertion axis into the inner surface and shows where it leaves more than 0.05 mm of material, with a colour map seen from the margin side. It is scored, and can hold or block the case (gate Seats on the preparation), only for the bur your lab declares — the bur typed here, or the one set in Settings › Milling machine. Without it the designer's MillingDiameter is used and the numbers are shown only. A bur larger than the one the design was compensated for (after the blank factor) is flagged.
  • Gingiva & hygiene (station 04): proximal contacts are measured as areas — also against the real neighbouring teeth in the scan (shown only until reviewed) — together with the open gingival embrasure ("black triangle") seen from the front, the pontic base shape and the connector-to-gingiva distance. Black triangles are scored only for anterior teeth and only when the gingiva comes from a gingiva design; a scanned gingiva may be displaced, so its values are shown, not scored.
  • Veneering space (station 06): when a tooth carries both its final shape and its framework (coping, framework or cutback layer), DentalQC measures the veneer thickness between them — thickest and thinnest veneer, how even it is, veneer under the cusp tips, the framework share where the material sets one, and framework sticking out of the final shape — with a colour map. Wall thickness and connectors are then measured on the framework, which is what gets milled. Case-level layers (Anatomic shapes + Reduced parts) are measured but not scored yet.
  • Settings › Lab materials: add the products you actually use from the catalogue (brand and line, under the seven generic classes), check each number against the manufacturer's IFU and press I have checked this against the IFU. Rules depend on context — region, restoration type, bridge length, pontics in a row, cantilevers — and every threshold carries its rule and source on the report. Until a product is confirmed it is measured but not scored; changing a number clears the confirmation. Products appear at the top of the material field on the measuring page; an exocad material code mapped to a product picks it for that tooth.
  • Settings › Milling machine: machine axes, smallest burs, the disc heights and blocks you stock, clearance and usable disc diameter. With discs or blocks listed, the gate Fits your blanks checks the height each unit or bridge needs, flat and at the best tilt (size only — sprues and nesting stay in CAM). With a 3-axis or 4-axis machine, the surface hidden from the blank axis is scored. Empty fields mean those checks stay “not checkable”.
  • Blank factor: zirconia only; use the number printed on the blank lot label (1.10–1.40), leave empty for the 1.25 default.
  • Patient (optional): Age (6–110), Sex, Intended tooth form (square / ovoid / tapering). The “Patient match” group is scored only when these are declared.

Left rail › Display layers

  • After measuring, toggle each overlay on the image: Wall thickness map, Occlusal clearance map, Crown outline, Symmetry axis, Per-tooth long axis, Zenith & incisal edge, Dental midline, Connector, Incisal curve & embrasures, FDI labels, mm grid.
  • Off by default: thickness map, clearance map and mm grid; the others are on.
  • Grid step slider: 0.5–5 mm, used by the mm grid layer.

Left rail › Annotation

  • The Annotation block only appears after measuring and starts folded: click the block title to open it. Once open, the drawing toolbar appears along the left edge of the image.
  • Show annotation layer checkbox: hide/show every mark without deleting.
  • Pin the correction list button: places numbered notes on the right teeth for the Corrections required items of this view.
  • Clear all button: deletes every annotation of the current view (asks first).
  • Mark list: click a row to select that mark on the image, the dot hides it, × deletes it.
  • Export image: Viewport PNG (image + visible layers + annotations, at 2×) · QC sheet PNG (with score, annotations and correction list) · Copy sheet to clipboard (paste straight into a chat or email).
  • Annotations belong to one arch and one view, live only in the browser, and are removed on re-measure (the projection is rebuilt). Ctrl+V pastes a screenshot onto this layer; dropping an image file on a measured viewport does the same.

Run panel (Measure button)

  • Once a file is chosen, the run panel appears in the middle of the viewport with the file name, a hint line and the run button. After measuring it moves up into the top strip so it does not cover the model.
  • The button label follows the state: Measure now (first time) · Measure new file (another file chosen) · Re-measure with new settings (settings changed since the last run) · Measure again (same settings).
  • Changing settings and re-measuring does not upload the file again — the server reuses the file it already has.
  • The hint line warns before measuring if only one STL file was chosen, or if the chosen files belong to several cases (use Batch then).
  • While measuring the viewport shows “Unpacking and measuring…”; a run usually takes a few seconds up to about 15 seconds.

Viewport

  • Top strip (after measuring): arch tabs Maxillary / Mandibular, view tabs Front · Occl. · R side · L side (plus 45° · Lingual · Insertion with the full view set), and the 3D button.
  • Zoom bar at the bottom right: − zoom out, current percentage, + zoom in, Fit, 1:1 (image pixel scale). The mouse wheel zooms too.
  • Status bar under the viewport: symmetry plane (yaw angle, number of tooth pairs, % match), px/mm, cursor position in mm, and run time.
  • Hover the image to see the floating note of the layer under the cursor.

Right column › QC board

  • The expand button ⤢ at the top (enabled after measuring) opens the full-screen Expanded QC board.
  • Blocks in order: Seven QC stations · Prescription cross-check (station 01) · Score by group · Intaglio surface (03) · Occlusal anatomy (06) · Corrections required · Interarch occlusion · Preparation & finish line (02) · Contact & connector (04) · Dynamic occlusion (05) · Patient match · Manufacturability & export (07) · Export gate · Cross-check 2D ↔ 3D.
  • Blocks without data read “No data yet.”; blocks with nothing out of tolerance read “Nothing exceeds tolerance.”.

Expanded QC board

  • A full-screen overlay that rearranges the same content as the right column into three zones: Verdict — is this case fit to leave (seven stations, prescription, gate), To fix, worst first, and Figures station by station.
  • The header shows the file name and score. Close with Close or the Esc key. Re-measuring while it is open updates it in place.

Measurement tables (bottom)

  • Ten tabs of raw figures: Per tooth · Antimeric pairs · Intaglio · Occlusal & profile · Prep & margin · Contact & connector · Incisal edge & embrasure · Occlusion · Manufacturability · Objects in file.
  • Drag the top edge to change the height; double-click the edge, click Collapse or press B to fold/unfold. With the edge focused, ↑ ↓ change the height. The height is remembered.

Active thresholds panel

  • Opened with the Your / Workshop thresholds button in the Scope block. Lists the metrics scored with custom thresholds: columns Metric, Default, In use, Why.
  • With no changes, it says plainly the case is scored with the default table. The run's tolerance profile still multiplies on top of these thresholds.
  • The footer reminds you that this table only changes scores; the export gate does not read it. Edit thresholds (if you may) or See the full table opens the Scoring thresholds page. Close with × or Esc.